Alpha II 111241 - Command List


List of Commands
Name Description Current Value Code Status Enabled Extension Error Level [f]
AAR Automatic accessory recognition X
ABP Absolute peak position 1000071 1000071 X
ACS Select Accessory 0 0 X
Enabled Call
0xFE80 (External Accessory)
Enabled Connected
ENTRY_NUM_ADDR=0xFE80
0xFEA0 (Front Accessory) X
Enabled Connected
ENTRY_NUM_ADDR=0xFEA0 X X
ADC Get Analog Input 0, 0
Enabled Call
0
Enabled Connected
DT_MOT=AdiAnalogIn
ADM Scanner adjustment mode Reinit 0 X
Enabled
0 (Reinit) X
1 (Fast Adjust Mode) X
2 (Stop) X
3 (Slow Adjust Mode) X
AQM Aquisition Mode Double Sided,Forward-Backward DD X
Enabled
DD (Double Sided,Forward-Backward) X
W
BMS Set beamsplitter KBr 1 X
Enabled Call
1 (KBr) X
Enabled Connected
DA_MOT=0xFE20 X X
ENTRY_NUM_ADDR=0xFE20 X X
BRK Break 0 0 X
Enabled
1 (Abort measurement) X
2 (Stop measurement) X
4 (Skip current "Waiting for Delay") X
8 (Skip current "Waiting for Trigger") X
16 (Skip current "Waiting for RDY") X
32 (Stop repeated measurement) X
CALIBRATE_SRC_DIG_POTI 0 0 X
CAN CAN Bus X
CCT Correct crystal temperature 0.000000 0.000000
Enabled Call
0
Enabled Connected
DT_MOT=TemperatureController
CHD Temperature control derivative gain 0.000000 0.000000
Enabled Call
0
Enabled Connected
DT_MOT=TemperatureController
CHI Temperature control integral gain 0.000000 0.000000
Enabled Call
0
Enabled Connected
DT_MOT=TemperatureController
CHN Measurement channel Sample Compartment 1 X
Enabled Call
1 (Sample Compartment) X
Enabled Connected
G_MOT=0xFF20, 0 X X
2 (Background)
Enabled Connected
G_MOT=0xFF20, 1
CHP Temperature control proportional gain 0.000000 0.000000
Enabled Call
0
Enabled Connected
DT_MOT=TemperatureController
CMA Correlation mask 0 0 X
Enabled Call
4 (ADC FS) X
8 (Vel) X
32 (IFG Length Diff) X
2048 (Unstable Sample Temperature)
Enabled Connected
DT_MOT=TemperatureController
8192 (Detector Not Ready) X
Enabled Connected
DT_MOT=Detector X X
CMD_STACK Command restore stack X
CNM Operator name X
COR Correlation on/off Off 0 X
Enabled
0 (Off) X
1 (On) X
CPD Derivative scanner control parameter 2000 2000 X
CPI Integral scanner control parameter 500 500 X
CPP Proportional scanner control parameter 5000 5000 X
CRM Correlation Reject Mask -1 -1 X
Enabled
14336 (Unstable Sample Temperature Or Vial Not Inserted Or Detector Not Ready) X
2048 (Unstable Sample Temperature) X
4096 (Vial Not Inserted) X
8192 (Detector Not Ready) X
DAC Set Analog Output 0, 0, 0
Enabled Call
0
Enabled Connected
DT_MOT=AdiAnalogOut
DAT_MOT Dummy command to enable/disable commands using units with given id and type 0 0
DA_MOT Unit ID 0 0
DEL Delay before measurement 0 0 X
DFO Data format DFO_RAW_IGRAM 1 X
Enabled
1 (DFO_RAW_IGRAM) X
DIA Universal command for diagnostics X
DIAG_CONNECTOR Set Diag Connector Test 0 0 X
Enabled
0x000000001 (Test: Single Pulse) X
0x000000010 (Pin: Err) X
0x000000020 (Pin: Fwd) X
0x000000040 (Pin: AcqEn) X
0x000000080 (Pin: TkDa) X
0x000100000 (Condition: On Static Error) X
0x000200000 (Condition: On Static Warning) X
0x000400000 (Condition: On Measurement Abort) X
DIS Display LED LED Off 0 X
Enabled
0 (LED Off) X
1 (LED Red) X
2 (LED Green) X
3 (LED Yellow) X
1000 (Restore LED Status) X
DISABLE_TCPIP_SETUP_MENU Prohibit Setting TCP/IP Parameters Via HTTP Server 0 0 X
DLY Stabilisation Delay 0 0 X
DTC Detector RT-DLATGS [Internal Pos.1] 65120 Ready X
Enabled Call
65088 ([External Pos.1])
Enabled Connected
G_MOT=0xFE40, 1
ENTRY_NUM_ADDR=0xFE40
65089 ([External Pos.2])
Enabled Connected
G_MOT=0xFE41, 1
ENTRY_NUM_ADDR=0xFE41
65120 (RT-DLATGS [Internal Pos.1]) X
Enabled Connected
G_MOT=0xFE60, 1 X X
ENTRY_NUM_ADDR=0xFE60 X X
65121 ([Internal Pos.2])
Enabled Connected
G_MOT=0xFE61, 1
ENTRY_NUM_ADDR=0xFE61
DT_MOT Dummy command to enable/disable commands units of given type
D_MINMAX Dummy command to assign unit min and max values to the command where this unit is included
D_MOT Unit ID 0 0, 0
ENTRY_NUM_ADDR Unit address used to find according entry number of a combobox command 0 0
ENTRY_NUM_ADDR_SAVING Unit address used to find according entry number of a combobox command (on low option) 0 0
FCM Fast Communication Mode Timeout, ms 0 0 X
FSS Full scan range 0 0 X
GCV Get Command Value X
GET_MOT_POS Get Motor Position 0 0 X
GET_POS Get Unit Position 0 0 X
GTP Get Sample Temperature 0
Enabled Call
0
Enabled Connected
DT_MOT=TemperatureController
GT_MOT Set motor to absolute position , 0 X
GUR Get Motor Resolution 0 0, 0 X
G_MOT Set motor to absolute position 65024 65024, 1 Ready X
G_MOT_FLOAT Set motor to absolute position using external integer parameter 0 0, 0.000000, 0.000000 X
G_MOT_INT Set motor to absolute position using external integer parameter 0 0, 0, 0 X
G_MOT_RANGE Set motor to absolute position; disable command if unit position is out of vaid range 0 0, 0 X
HFW Wanted high frequency limit 8000.000000 8000.000000 X
HMR Instrument humidity range 0.000000 0.000000 X
HMT Instrument humidity target 0.000000 0.000000 X
HPF High Pass Filter On 1 X
Enabled
0 (Off) X
1 (On) X
ITC Instrument test class 0 0 Ready X
ITI Instrument test interval 0 0 X
ITO OPUS inactivity timeout interval, s 20 20 X
LFT Lifetime interval 0 0 X
LFW Wanted low frequency limit 0.000000 0.000000 X
LGL Log Level Bitmask Standard 1 X
Enabled
1 (Standard) X
2 (Verbose) X
0x00F0 (Data Acquisition) X
0x0F00 (CAN) X
LSR Switch laser on/off 1 1 X
LWN Laser wave number 0.000000 0.000000 X
MIN Scan Time in Minutes 0.000000 0.000000 X
MOH Set motor to hold position 0 0, 0 X
MOR Set motor to relative position 0 0, 0 X
MOR_INT Set motor to relative position using external integer parameter 0 0, 0 X
MOS Set motor to home position 0 0, 0 X
MOT Set motor to absolute position 0 0, 0 X
MOT_INT Set motor to absolute position using external integer parameter 0 0, 0 X
MOT_STAT Wait for device status update 0 0 X
MSR Motor status register 0 0, 0 X
NSS Number of sample scans 4 4 X
OPF Optical Filter Wheel Open 0 Ready X
Enabled Call
0 (Open) X
Enabled Connected
G_MOT=0xFF00, 0 X X
1 (Filter Polystyrene) X
Enabled Connected
G_MOT=0xFF00, 1 X X
2 (Poly50) X
Enabled Connected
G_MOT=0xFF00, 2 X X
OVS Overscan range 2000 2000 Ready X
PBF Set Pass Band Factor 0.430000 0.430000 X
PGN Preamplifier gain Standard 1 X
Enabled
1 (Standard) X
RBT Read Digital Input 0, 0
Enabled Call
0
Enabled Connected
DT_MOT=AdiDigitalIn
RDR Sampling Data Rate *?RDS=4096?* 21600 X
Enabled
21600 (*?RDS=4096?*) X
RDX Instrument ready mask Source 32 X
Enabled
2 (Detector) X
8 (Sample Temperature) X
32 (Source) X
34 (OVP) X
RDY Instrument ready status 0 0 X
REP Repeat measurement count 0 0 X
REP_CMD Send command with current value X
RES Resolution 14.220000 14.220000 X
RSN Running sample number 0 0 X
SBT Set Digital Output 0, 0, 0
Enabled Call
0
Enabled Connected
DT_MOT=AdiDigitalOut
SBY Probe Trigger on/off Standby Off 0 X
Enabled
0 (Standby Off) X
1 (Standby On) X
2 (Standby On OPUS Timeout)
SET_CAN_RECEIVE_BUFFER_SIZE_MESSAGES Set CAN receive buffer size, messages 512 512 X
SET_DACQ_MEM_POOL_SIZE_BUFFERS Set data acquisition memory pool size, buffers 6 6 X
SET_DACQ_MEM_POOL_SIZE_BYTES Set data acquisition memory pool size, bytes 4194304 4194304 X
SET_HTTP_SERVER_TX_POOL_SIZE_BUFFERS Set data http server tx memory pool size, buffers 4 4 X
SET_OUTPUT Set Output of Temperature Controller, percentage of maximal power 0.000000 0.000000 X
SET_SCR_REINIT_ON 1 1 X
SET_TEST_DOWNLOAD_FILE_SIZE_BYTES Set test download file size, bytes 524288 524288 X
SFM Sample Form X
SNM Sample Name X
SON Probe Trigger on/off Off 0 X
Enabled
0 (Off) X
1 (Internal Trigger) X
2 (External Trigger) X
3 (Both Trigger) X
SOT Probe Trigger on/off Scans 0 X
Enabled Call
0 (Scans) X
1 (Minutes) X
2 (Rotations)
Enabled Connected
DT_MOT=Rotator
SRC Source MIR 101 Ready X
Enabled Call
0 (Off) X
Enabled Connected
G_MOT=0xFE00, 0 X X
101 (MIR) X
Enabled Connected
G_MOT=0xFE00, 1 X X
ENTRY_NUM_ADDR=0xFE00 X X
SST Set Sample Temperature 0.000000 0.000000
Enabled Call
0
Enabled Connected
DT_MOT=TemperatureController
D_MINMAX=TemperatureController
STS Set Instrument Test Status Disable 0 X
Enabled Call
0 (Disable) X
Enabled
ITI=0 X
1 (Reinitialize) X
Enabled
ITI=-1 X
2 (Set To Failed) X
Enabled
ITI=-2 X
5 (Increase Interval) X
Enabled
ITI=-5 X
6 (Remove) X
Enabled
ITI=-6 X
7 (Set To Expired) X
Enabled
ITI=-7 X
TEST_OUTP_STEP Test Command 0 0 X
TMR Instrument temperature range 0.000000 0.000000 X
TMT Instrument temperature target -300.000000 -300.000000 X
TRQ Turn frequency 0 0 X
TRW Input signal selection IR 0 X
Enabled
0 (IR) X
1 (Test ramp) X
2 (Shorted input) X
3 (Laser) X
TSR Tolerance scan range 2000 2000 X
UWN ITC / channel specific LWN On/Off On 1 X
Enabled
0 (Off) X
1 (On) X
VEL Velocity 7.500000 7.500000 X
Enabled
7.5 (7.5 KHz) X
VSN Firmware version X
WAIT_FOR_UNIT_READY Wait for unit ready 0 0, 0 X
WAIT_MS Wait for delay, ms 0.000000 0.000000 X
WRK Set type of command execution 0 0 X
XND Additional delay of the IR signal relative to the laser one [ns] 0.000000 0.000000 X
XWN Instrument test class specific laser wave number X
_GR Get measurement result 0 0 X F
_ME Measure 0 0 X F
_SP Set measurement parameters 0 0 X
Messages
Timestamp Code Level Sender/Answer Command Message
02:00:12 60334 A ITC 0 ITC=0 Instrument Test Disabled
02:01:06 40650 U SourceUnit Expected lifetime elapsed, unreliable operation End of lifetime is nearly reached, spare part will be required

Page generated on Fri, 02 Jan 1970 02:41:32